1) Turn on the display. The screen will count down from [999999] and then show [0.00]. If [0.00] does not appear, press the [ZERO] key to reset to zero.
2) Use tweezers to place the sample on the measuring platform.
3) Press [ENTER] when the stable indicator appears. The screen will show [SAVE-A], followed by [------], indicating that the weight of the sample in air is being recorded. Once the memory process is complete, the pointer will move to M, representing the weight in air.
4) Carefully pick up the sample with tweezers, rinse it in alcohol, and place it on the hanging bar in water. When the stable symbol appears again, press [ENTER]. The screen will show [****], then [SAVE-B], confirming that the weight in water is being stored. After the process completes, the pointer will point to SG, displaying the specific gravity value.
5) After recording the required data, press [ENTER] to return to the test mode and proceed to test the next sample.
Additional Options:
After the specific gravity value is displayed, press the [B] key to view the following options:
SHO-SG: Specific Gravity Value
SHO-V: Volume Value
SHO-%: Percentage Value
- - Water Weight Value
This device provides a precise and efficient way to measure the specific gravity of samples. It's ideal for laboratory settings or quality control processes where accuracy is essential. Make sure to follow each step carefully to ensure correct readings and avoid errors. Regular maintenance and calibration of the equipment are also recommended to maintain performance over time.